REFERENCE TABLES

Table 6-1 provides detail information for main menu items.
Table 6-2 provides detail information for sub-menu items.


Main Menu Items Table 6-1
Main Menu Display
Sub-Menu Display
Sub-Menu Information
MENU Key
 
Use ADV and SET key to perform changes
"Func"
Function
Select rate mode (refer to Section 4) or totalize mode (refer to Section 5).
"SCALE"
Scale Factor
Select scale factor as a divider or multiplier. Enter scale factor value from -.9.9.9.9.9 through 9.9.9.9.9.9. After measuring the input (totalling), the meter multiplies or divides the result by the scale factor then adds the offset.
"Offset"
"Offset Value"
Select offset value from -.9.9.9.9.9 through 9.9.9.9.9.9. In the totalize mode, the meter calculates and shows the offset value after you press the RESET button. The meter will then be at a starting point for totalizing.
"DEC PT"
Decimal Point
Select where decimal point shows in the digital display. You may set a decimal point at any of six available positions. You may also choose a floating decimal point (auto-range).
"Sp LO"

"SP HI"
Setpoints
Shows only if the relay option board is installed. Similar to offset, -.9.9.9.9.9 to 9.9.9.9.9.9. shows. After applying the scale factor and offset, the displayed measurement is compared to the setpoint values.
"CONFIG"
Configuration
Program internal software switches to specify baud rate, activate external or internal gate input, and modify alarm conditions. Refer to Table 6-3 for configuration detail.
"AN LO"

"AN HI"
Analog Low


Analog High
Display value for low analog output.

Display value for high analog output.
Note: Install S2-B to unlock this prompt.
"noStor"
NoStore/Store
Select "noStor" to save to the volatile (active) memory. Select "Store" to save to the nonvolatile (permanent) memory. Nonvolatile memory is not effected by power loss.

Sub Menu Items Table 6-2
Main Menu Display
Sub-Menu Display
Sub-Menu Information
MENU Key
 
Use ADV and SET key to perform changes
"Func"
"rate"
In the "rAtE" mode, the meter acts as a ratemeter for measuring the rate at which counts are received. The meter shows in Hz (pulse/sec), and may also show in kHz, RPM (for tachometer applications), feet/sec and other engineering units.
"total"
In the "totAL" mode, the meter acts as a totalizer for registering a precise total count from mechanical, photoelectric, electromagnetic or electronic inputs or detectors. Use for cumulative totals and stopwatch applications. The meter up-counts (positive scale factor) and may be configured to down-count (negative scale factor).
"SCALE"
"IN/SC"

"IN.SC"






Scale factor is a divider.

Scale factor is a multiplier.

Multiply or divide input by any desired scale factor from -99999 to 999999. After measuring the input, the meter multiplies or divides the result by the scale factor and then adds the offset.
"DEC Pt"
"F.FFFFF"
Decimal point moves from left to right through the digital display. You may set a fixed decimal point at any of the six available positions (e.g. -.9.9.9.9.9 to 9.9.9.9.9.9.). This mode is useful for low frequency, low resolution pulse inputs, such as from paddle-wheel sensors. For example, to have the meter show to the nearest 1/100 of a gallon, select the decimal point "FFFF.FF". When you set a fixed decimal point, the meter maintains a meaningful unit of measurement. For example, if the input frequency is 10 Hz, the function is rate and the fixed decimal point equals 3, the display reads 10.00, not .10.
You may also choose a floating decimal point (auto-range). Auto-range automatically adjusts the decimal point location to provide the highest measurement resolution possible. This mode is useful for high frequency, high resolution pulse inputs, such as outputs from turbine meters or low flow sensors.

The 6 digits of Configuration show as a sub-menu of CONFIG.
Meter Configuration Detail Table 6-3
(Factory defaults are bold)

6 Digits of Configuration
0
0
0
0
0
0

0
1

Determines analog output signal.
4-20 mA
0-10 V
0
1

Unlatched
Latched
0
1

Setpoint LO works as active below the setpoint
Setpoint LO works as active above the setpoint (similar to setpoint HI).
0
1

Message handshake - If RTS input is true, sends complete message
Character handshake - Checks RTS input before sending each character, and sends characters only while RTS is true.

0
1
Determines the communication rate.
9600 baud rate
1200 baud rate
0
1
Transmits alarm status (total of 9 characters)
Does not transmit alarm status character
Help Price List Specifications Print Manual Manuals Home
 Home Home  Table of Contents TOC  Back Back  Next Next